We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVE: This study aimed at measuring the magnetic field strength of commonly used types of audiological testing equipment and determine their effects on the three most commonly used programmable ventriculoperitoneal (VP) shunts to try and quantify the risk of a VP shunt being reprogrammed during audiological testing. DESIGN: In this in vitro study, magnetic field strength was measured for TDH 39 supra-aural earphones, B71 bone vibrator, ALGO 3i probe, Bio-logic Nav Pro probe, Otodynamics otoport insert earphone and Madsen Zodiac tympanometry probe. STUDY SAMPLE: Magnetic field strength associated with transducers placed on a model of a skulls having implanted Miethke ProGAV 2.0, Medtronic Strata II and Codman Hakim programmable VP shunts was measured. RESULTS: The supra-aural earphones had a magnetic field strength of 14 mT at 0 mm, which dropped to 0 mT at 10 mm away from the transducer. All other equipment had a magnetic field strength of 3.5 mT or less at 0 mm. There was no instance of reprogramming of the shunts by the transducers. CONCLUSIONS: The findings suggest that the risk of inadvertent valve-reprogramming by the transducers is extremely small. However, care should be taken to avoid placing any of the transducers directly over the shunt.

More information Original publication

DOI

10.1080/14992027.2021.1881174

Type

Journal article

Publication Date

2022-02-01T00:00:00+00:00

Volume

61

Pages

173 - 176

Total pages

3

Keywords

Audiology assessment, audiology equipment, magnetic strength, newborn hearing screening, programmable ventriculoperitoneal shunts, supra-aural headphones, Acoustic Impedance Tests, Equipment Design, Humans, Transducers, Ventriculoperitoneal Shunt