We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVE: To assess risk factors and predictors of failure of the Hem-o-lok(TM) vascular clip (Weck Closure Systems, Research Triangle Park, NC, USA) using vessels harvested from a porcine model. MATERIALS AND METHODS: Vessels of various diameters were harvested from a porcine model, clipped at 90 degrees or 45 degrees using the Hem-o-lok clip and then cut either flush or with a 1-mm cuff. The vessels were then connected to a burst-pressure device and pressures required to burst the clip or to cause it to leak were measured. RESULTS: The Hem-o-lok clip leaked or burst when the vessel to which it was applied was cut flush. The clip became even more likely to fail if the angle of application of the clip was not at 90 degrees to the vessel surface. CONCLUSION: The Hem-o-lok vascular clip is safe if it is applied at 90 degrees to the vessel surface and, more importantly, if a 1-mm cuff is left between the clip and the point at which the vessel is divided. We would therefore discourage the practice of not leaving this cuff of tissue, in an attempt to maximize vessel length during laparoscopic donor nephrectomy.

Original publication

DOI

10.1111/j.1464-410X.2008.08211.x

Type

Journal article

Journal

BJU Int

Publication Date

05/2009

Volume

103

Pages

1410 - 1412

Keywords

Animals, Blood Loss, Surgical, Equipment Failure, Hemostasis, Surgical, Humans, Kidney, Laparoscopy, Nephrectomy, Risk Factors, Surgical Instruments, Swine